<?xml version='1.0'?>
<!DOCTYPE art SYSTEM 'http://www.biomedcentral.com/xml/article.dtd'>
<art>
<ui>1746-1596-7-30</ui>
<ji>1746-1596</ji>
<fm>
<dochead>Research</dochead>
<bibl>
<title><p>Comparison of HemoCue<sup>&#174; </sup>hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan</p></title>
<aug>
<au id="A1" ca="yes"><snm>Adam</snm><fnm>Ishag</fnm><insr iid="I1"/><email>ishagadam@hotmail.com</email></au>
<au id="A2"><snm>Ahmed</snm><fnm>Samah</fnm><insr iid="I1"/><email>samah289@gmail.com</email></au>
<au id="A3"><snm>Mahmoud</snm><mi>H</mi><fnm>Mahmoud</fnm><insr iid="I1"/><email>mheljaaly2020@hotmail.com</email></au>
<au id="A4"><snm>Yassin</snm><mi>I</mi><fnm>Mohammed</fnm><insr iid="I1"/><email>dr_yassinibrahim@yahoo.com</email></au>
</aug>
<insg>
<ins id="I1"><p>Faculty of Medicine, University of Khartoum, P.O. Box 102, Khartoum, Sudan</p></ins>
</insg>
<source>Diagnostic Pathology</source>
<issn>1746-1596</issn>
<pubdate>2012</pubdate>
<volume>7</volume>
<issue>1</issue>
<fpage>30</fpage>
<url>http://www.diagnosticpathology.org/content/7/1/30</url>
<xrefbib><pubidlist><pubid idtype="doi">10.1186/1746-1596-7-30</pubid><pubid idtype="pmpid">22436620</pubid></pubidlist></xrefbib></bibl>
<history><rec><date><day>4</day><month>2</month><year>2012</year></date></rec><acc><date><day>21</day><month>3</month><year>2012</year></date></acc><pub><date><day>21</day><month>3</month><year>2012</year></date></pub></history><cpyrt><year>2012</year><collab>Adam et al; licensee BioMed Central Ltd.</collab><note>This is an Open Access article distributed under the terms of the Creative Commons Attribution License (<url>http://creativecommons.org/licenses/by/2.0</url>), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.</note></cpyrt>
<abs>
<sec><st><p>Abstract</p></st>
<sec><st><p>Background</p></st>
<p>Assessment of hemoglobin is one of the most reliable indicators for anemia, and is widely used to screen for anemia among pregnant women. The HemoCue<sup>&#174; </sup>has been widely used for as a point-of-care device for hemoglobin estimation in health facilities. Previous studies showed contradictory results regarding the accuracy of HemoCue<sup>&#174;</sup>.</p>
</sec>
<sec><st><p>Methods</p></st>
<p>This was a hospital-based cross sectional study carried- out among pregnant women at Khartoum hospital in Sudan to find out whether the measurement of hemoglobin concentration by HemoCue<sup>&#174; </sup>using venous or capillary samples was comparable to that of the automated hematology analyzer as standard. Bland and Altman method was used to compare the measurements with an acceptable difference of &#177; 1.0 g/dl.</p>
</sec>
<sec><st><p>Results</p></st>
<p>Among the 108 subjects in this study the mean (SD) level of hemoglobin level using HemoCue<sup>&#174; </sup>venous sample, HemoCue<sup>&#174; </sup>capillary sample and automated hematology analyzer were 12.70 (1.77), 12.87 (2.04) and 11.53 (1.63) g/dl, respectively. Although the correlations between the measurements were all significant there was no agreement between HemoCue<sup>&#174; </sup>and automated hematology analyzer. The bias + SD (limits of agreement) for HemoCue<sup>&#174; </sup>venous versus hematology analyzer was 1.17 &#177; 1.57 (-1.97, 4.31) g/dl, HemoCue<sup>&#174; </sup>capillary versus hematology analyzer was 1.34 &#177; 1.85 (-2.36, 5.04) g/dl, and HemoCue<sup>&#174; </sup>venous versus HemoCue<sup>&#174; </sup>capillary samples was 017 &#177; 1.90 and (3.97-3.63) g/dl.</p>
</sec>
<sec><st><p>Conclusion</p></st>
<p>Hemoglobin concentration assessment by HemoCue<sup>&#174; </sup>using either venous or capillary blood samples has shown unacceptable agreement with automated hematology analyzer.</p>
</sec>
<sec><st><p>Virtual slides</p></st>
<p>The virtual slide(s) for this article can be found here: <url>http://www.diagnosticpathology.diagnomx.eu/vs/8797022296725036</url></p>
</sec>
</sec>
</abs>
</fm>
<bdy>
<sec><st><p>Introduction</p></st>
<p>Anemia is one of the most important causes of morbidity and mortality in developing countries, especially among pregnant women <abbrgrp><abbr bid="B1">1</abbr></abbrgrp>. The world prevalence of anemia in pregnant women and non-pregnant women is 41.8% and 30.2% respectively <abbrgrp><abbr bid="B2">2</abbr></abbrgrp>. Pregnant Sudanese women are susceptible to anemia regardless to their age and parity and anemia is one of the leading causes of maternal and perinatal morbidity and mortality <abbrgrp><abbr bid="B3">3</abbr><abbr bid="B4">4</abbr><abbr bid="B5">5</abbr></abbrgrp>.</p>
<p>Assessment of hemoglobin is one of the most reliable indicators for anemia, and is widely used to screen for anemic individuals, and to evaluate responses to interventions <abbrgrp><abbr bid="B6">6</abbr></abbrgrp>. Hemoglobin concentration is routinely measured using automated hematology analyzers. Although, these are very accurate and reliable, they are expensive and problems of samples transport to the laboratory may delay treatment resulting in preventable deaths <abbrgrp><abbr bid="B7">7</abbr></abbrgrp>. In poor resources where automated hematology analyzer is not available other methods of low price and that require less skill are highly needed. Although Cyanmethemoglobin method is cheaper and often used, it takes more time. The semi-quantitative gravimetric copper sulfate method which is used in blood donation is very easy and inexpensive; it does not provide an acceptable degree of accuracy.</p>
<p>The HemoCue<sup>&#174; </sup>hemoglobin photometer has been widely used for as a point-of-care device for hemoglobin estimation in mobile blood donations and critical care areas in health facilities <abbrgrp><abbr bid="B8">8</abbr></abbrgrp>. Previous studies showed contradictory results regarding the accuracy of HemoCue<sup>&#174;</sup>; some of research reported a high accuracy of HemoCue<sup>&#174; </sup>compared with standard laboratory methods <abbrgrp><abbr bid="B9">9</abbr><abbr bid="B10">10</abbr></abbrgrp>. However, others did not recommend this device in general practice <abbrgrp><abbr bid="B11">11</abbr><abbr bid="B12">12</abbr><abbr bid="B13">13</abbr></abbrgrp>. Few published data exist on the accuracy of HemoCue<sup>&#174; </sup>in the measurement of hemoglobin during pregnancy <abbrgrp><abbr bid="B14">14</abbr><abbr bid="B15">15</abbr></abbrgrp>. The aim of this study was to compare the accuracy of HemoCue<sup>&#174; </sup>using venous and capillary samples with that of the automated hematology analyzer in the measurement of hemoglobin among pregnant Sudanese women at Khartoum Hospital.</p>
</sec>
<sec><st><p>Material and methods</p></st>
<p>This was a cross sectional study carried- out among pregnant women at Khartoum hospital in Sudan during the period of October through December 2011. Convenient sampling method was used in the study in which all available subjects who fulfilled the inclusion criteria at the Khartoum hospital antenatal care clinic were recruited in the study every day until total number of sample size was achieved. Sample size of 108 subjects was calculated based on a 2-sided hypothesis tests using Epiinfo with 80% power and confidence interval of 95%. After signing an informed consent venous and capillary blood samples were taken from each woman to be measured by HemoCue<sup>&#174; </sup>(venous and capillary) and the automated analyzer.</p>
<p>Capillary blood samples were collected by finger prick in the middle finger of left hand, after cleaning and massaging the finger to facilitate blood flow. Venous blood samples were collected into vacutainer tubes and were analyzed immediately by HemoCue<sup>&#174; </sup>and the tubes were sent to the medical laboratory of the hospital for the analyses by automated analyzer.</p>
<sec><st><p>HemoCue<sup>&#174; </sup>portable photometer</p></st>
<p>The HemoCue<sup>&#174; </sup>B-Hemoglobin system (HemoCue<sup>&#174; </sup>AB, &#196;ngelholm, Sweden) consists of disposable microcuvettes containing reagent in a dry form and a single purpose designed photometer. The microcuvettes were stored in a dry place at room temperature. Once opened, they were tightly closed and stored at the same conditions to maintain their integrity and shelf life. The reaction in the microcuvette is a modified azide-methemoglobin reaction. Sodium deoxycholate haemolyses erythrocytes and hemoglobin is released. Sodium nitrite converts hemoglobin to methemoglobin which, together with sodium azide, gives azidemethemoglobin. The absorbance is measured at two wavelengths (570 nm and 880 nm) in order to compensate for turbidity in the sample. The test was performed as stated by the manufacturer <abbrgrp><abbr bid="B16">16</abbr></abbrgrp>.</p>
</sec>
<sec><st><p>Automated hematology analyzer</p></st>
<p>The Sysmex KX21N (Sysmex Corporation, Kobe, Japan) is an automated blood cell counter intended for in vitro diagnostic use in clinical laboratories. It is a compact, fully automated hematology analyzer with simultaneous analysis of 18 parameters in whole blood mode and capillary blood mode. It measures the hemoglobin concentration using a non-cyanide hemoglobin method (STROMATOLYSER WH). The instrument has been proven to provide accurate and reliable results including hemoglobin concentrations <abbrgrp><abbr bid="B17">17</abbr><abbr bid="B18">18</abbr></abbrgrp>. The test was performed as stated in the manufacturer's manual <abbrgrp><abbr bid="B19">19</abbr></abbrgrp>.</p>
</sec>
<sec><st><p>Quality control</p></st>
<p>The HemoCue<sup>&#174; </sup>photometer was checked on a daily basis using the control cuvette and a standard of known concentration. A three set controls were run daily to ensure the function of the Sysmex.</p>
</sec>
<sec><st><p>Statistics</p></st>
<p>All statistical analyses were performed using SPSS software (SPSS Inc., Chicago, IL, USA, version 16.0). Hemoglobin levels were measured using two instruments; HemoCue<sup>&#174; </sup>and automated hematology analyzer. HemoCue<sup>&#174; </sup>used two types of samples; venous as well as capillary samples. Pairs of hemoglobin measurements were compared as follows: HemoCue<sup>&#174; </sup>venous sample versus automated hematology analyzer, HemoCue<sup>&#174; </sup>capillary sample versus automated hematology analyzer, HemoCue<sup>&#174; </sup>venous sample versus HemoCue<sup>&#174; </sup>capillary sample. Pearson Correlation analysis was performed and correlation coefficient (r) was calculated. The Mean of differences (bias), standard deviation of differences (SD), and limits of agreement (Mean &#177; 2 &#215; SD) were calculated according to the Bland and Altman method <abbrgrp><abbr bid="B20">20</abbr></abbrgrp>. Limits of agreement not exceeding &#177;1 g/dl between any two pairs of methods were considered to be clinically acceptable.</p>
</sec>
<sec><st><p>Ethics</p></st>
<p>This study was completely conducted voluntarily and all respondents were given brief explanations about the purpose and procedures of the study to be used. Voluntarily consent was signed by each subject before taking the blood. Ethical clearance was obtained from the institute Board.</p>
</sec>
</sec>
<sec><st><p>Results</p></st>
<p>A total number of 108 pregnant women were recruited in this study. Their mean (SD) of the age and gestational age were 27.6 (6.8) years and 24.9 (10.2) weeks.</p>
<p>The mean (SD) hemoglobin level was 12.70 (1.77), 12.87 (2.04) and 11.53 (1.63) g/dl using HemoCue<sup>&#174; </sup>venous sample, HemoCue<sup>&#174; </sup>capillary sample and automated hematology analyzer, respectively (Table <tblr tid="T1">1</tblr>).</p>
<tbl id="T1"><title><p>Table 1</p></title><caption><p>Hemoglobin level measured using HemoCue<sup>&#174; </sup>venous sample, HemoCue<sup>&#174; </sup>capillary sample, and automated hematology analyzer (g/dl)</p></caption><tblbdy cols="4">
      <r>
         <c ca="left">
            <p>
               <b>Method of measurement</b>
            </p>
         </c>
         <c ca="left">
            <p>
               <b>Mean &#177; SD</b>
            </p>
         </c>
         <c ca="left">
            <p>
               <b>Median</b>
            </p>
         </c>
         <c ca="left">
            <p>
               <b>(min; max)</b>
            </p>
         </c>
      </r>
      <r>
         <c cspan="4">
            <hr/>
         </c>
      </r>
      <r>
         <c ca="left">
            <p>HemoCue<sup>&#174; </sup>(venous)</p>
         </c>
         <c ca="left">
            <p>12.70 &#177; 1.77</p>
         </c>
         <c ca="left">
            <p>12.80</p>
         </c>
         <c ca="left">
            <p>(8.90; 17.90)</p>
         </c>
      </r>
      <r>
         <c ca="left">
            <p>HemoCue<sup>&#174; </sup>(capillary)</p>
         </c>
         <c ca="left">
            <p>12.87 &#177; 2.04</p>
         </c>
         <c ca="left">
            <p>12.90</p>
         </c>
         <c ca="left">
            <p>(8.20; 17.70)</p>
         </c>
      </r>
      <r>
         <c ca="left">
            <p>Hematology analyzer</p>
         </c>
         <c ca="left">
            <p>11.53 &#177; 1.63</p>
         </c>
         <c ca="left">
            <p>11.40</p>
         </c>
         <c ca="left">
            <p>(8.10; 15.00)</p>
         </c>
      </r>
   </tblbdy></tbl>
<sec><st><p>HemoCue<sup>&#174; </sup>venous versus automated hematology analyzer</p></st>
<p>There was a positive correlation (<it>r </it>= 0.58, <it>P </it>&lt; 0.001) between hemoglobin levels by using HemoCue<sup>&#174; </sup>venous sample versus automated hematology analyzer. The mean difference with limits of agreement between the two reading was 1.17 (-1.97, 4.31) g/dl (Table <tblr tid="T2">2</tblr> and Figure <figr fid="F1">1</figr>).</p>
<tbl id="T2"><title><p>Table 2</p></title><caption><p>Correlation, bias, and limits of agreement between hemoglobin level using HemoCue<sup>&#174; </sup>and automated hematology analyzer</p></caption><tblbdy cols="4">
      <r>
         <c ca="left">
            <p>
               <b>Comparison of methods</b>
            </p>
         </c>
         <c ca="left">
            <p>
               <b>Correlation Coefficient</b>
            </p>
         </c>
         <c ca="left">
            <p>
               <b>Bias &#177; SD (95%CI)</b>
            </p>
         </c>
         <c ca="left">
            <p>
               <b>Limits of agreement</b>
            </p>
         </c>
      </r>
      <r>
         <c cspan="4">
            <hr/>
         </c>
      </r>
      <r>
         <c ca="left">
            <p>HemoCue<sup>&#174; </sup>(venous) vs. Hematology analyzer</p>
         </c>
         <c ca="left">
            <p>0.58</p>
         </c>
         <c ca="left">
            <p>1.17 &#177; 1.57 (0.87; 1.47)</p>
         </c>
         <c ca="left">
            <p>-1.97 to 4.31</p>
         </c>
      </r>
      <r>
         <c ca="left">
            <p>HemoCue<sup>&#174; </sup>(capillary) vs. Hematology analyzer</p>
         </c>
         <c ca="left">
            <p>0.51</p>
         </c>
         <c ca="left">
            <p>1.34 &#177; 1.85 (0.99; 1.69)</p>
         </c>
         <c ca="left">
            <p>-2.36 to 5.04</p>
         </c>
      </r>
      <r>
         <c ca="left">
            <p>HemoCue<sup>&#174; </sup>(venous) vs. HemoCue<sup>&#174; </sup>(capillary)</p>
         </c>
         <c ca="left">
            <p>0.51</p>
         </c>
         <c ca="left">
            <p>-0.17 &#177; 1.90 (-0.53; 0.19)</p>
         </c>
         <c ca="left">
            <p>-3.97 to 3.63</p>
         </c>
      </r>
   </tblbdy></tbl>
<fig id="F1"><title><p>Figure 1</p></title><caption><p>Bland and Altman plot for hemoglobin level by using HemoCue<sup>&#174; </sup>venous sample versus automated hematology analyzer (g/dl)</p></caption><text>
   <p><b>Bland and Altman plot for hemoglobin level by using HemoCue<sup>&#174; </sup>venous sample versus automated hematology analyzer (g/dl)</b>.</p>
</text><graphic file="1746-1596-7-30-1" hint_layout="double"/></fig>
</sec>
<sec><st><p>HemoCue<sup>&#174; </sup>capillary versus automated hematology analyzer</p></st>
<p>There was a positive correlation between hemoglobin levels by using HemoCue<sup>&#174; </sup>capillary sample versus automated hematology analyzer (<it>r </it>= 0.51, <it>P </it>&lt; 0.001). The mean difference with limits of agreement between hemoglobin levels by using HemoCue<sup>&#174; </sup>capillary sample versus automated hematology analyzer was 1.34 (-2.36, 5.04) g/dl (Table <tblr tid="T2">2</tblr> and Figure <figr fid="F2">2</figr>).</p>
<fig id="F2"><title><p>Figure 2</p></title><caption><p>Bland and Altman plot for hemoglobin level by using HemoCue<sup>&#174; </sup>capillary sample versus automated hematology analyzer (g/dl)</p></caption><text>
   <p><b>Bland and Altman plot for hemoglobin level by using HemoCue<sup>&#174; </sup>capillary sample versus automated hematology analyzer (g/dl)</b>.</p>
</text><graphic file="1746-1596-7-30-2" hint_layout="double"/></fig>
</sec>
<sec><st><p>HemoCue<sup>&#174; </sup>venous versus HemoCue<sup>&#174; </sup>capillary sample</p></st>
<p>There were positive correlation between hemoglobin levels by using HemoCue<sup>&#174; </sup>venous sample versus HemoCue<sup>&#174; </sup>capillary sample (<it>r </it>= 0.51, <it>P </it>&lt; 0.001). The mean difference with limits of agreement between hemoglobin levels by using HemoCue<sup>&#174; </sup>venous sample versus HemoCue<sup>&#174; </sup>capillary sample was -0.17 (-3.97, 3.63) g/dl (Table <tblr tid="T2">2</tblr> and Figure <figr fid="F3">3</figr>).</p>
<fig id="F3"><title><p>Figure 3</p></title><caption><p>Bland and Altman plot for hemoglobin level by using HemoCue<sup>&#174; </sup>venous sample versus HemoCue<sup>&#174; </sup>capillary sample (g/dl)</p></caption><text>
   <p><b>Bland and Altman plot for hemoglobin level by using HemoCue<sup>&#174; </sup>venous sample versus HemoCue<sup>&#174; </sup>capillary sample (g/dl)</b>.</p>
</text><graphic file="1746-1596-7-30-3" hint_layout="double"/></fig>
<p>According to the previously pre-defined clinical acceptable limits of &#177; 1 g/dl, the 2 methods could not be considered as interchangeable.</p>
</sec>
</sec>
<sec><st><p>Discussion</p></st>
<p>Recently HemoCue<sup>&#174; </sup>portable hemoglobin photometer using venous or capillary blood samples has been widely used for quick assessment of hemoglobin concentrations; especially in poor settings where skills and resources are limited.</p>
<p>The accuracy of HemoCue<sup>&#174; </sup>for measuring hemoglobin in clinical settings is still a matter of controversy. The results of this study have shown that the hemoglobin concentration of HemoCue<sup>&#174; </sup>using either venous or capillary blood samples have lower level of precision and was not comparable with that of automated hematology analyzer; the limits of agreement were larger than the predefined clinically acceptable limits of &#177;1 g/dl. This is goes with the previous findings of the studies conducted among pregnant women at high altitude <abbrgrp><abbr bid="B14">14</abbr></abbrgrp> and adult patients hospitalized in surgical intensive care unit <abbrgrp><abbr bid="B21">21</abbr></abbrgrp>. The difference between the readings has been explained by the use of only one microcuvette with HemoCue<sup>&#174; </sup>especially when using capillary blood samples and laboratory values, and advised loading multiple microcuvettes and averaging the hemoglobin values obtained has been proposed <abbrgrp><abbr bid="B22">22</abbr></abbrgrp>.</p>
<p>However, other studies showed that the results obtained by using HemoCue<sup>&#174; </sup>for hemoglobin assessment among pregnant women were comparable to that of automated hematology analyzer as a standard <abbrgrp><abbr bid="B9">9</abbr></abbrgrp>. Bernard <it>et al.</it>, found that the results of hemoglobin concentration among pregnant and non-pregnant populations using HemoCue<sup>&#174; </sup>were comparable to that of automated hematology analyzer and Cyanmethemoglobin methods <abbrgrp><abbr bid="B23">23</abbr></abbrgrp>. Other studies which were conducted in different settings and populations such as patients with gastrointestinal bleeding, surgical patients repeated measurement of one sample, urban general practice, neonates, patients undergoing aortic surgery in the theatre and blood donors recommended HemoCue<sup>&#174; </sup>for the hemoglobin estimation <abbrgrp><abbr bid="B8">8</abbr><abbr bid="B24">24</abbr><abbr bid="B25">25</abbr><abbr bid="B26">26</abbr><abbr bid="B27">27</abbr><abbr bid="B28">28</abbr><abbr bid="B29">29</abbr><abbr bid="B30">30</abbr><abbr bid="B31">31</abbr></abbrgrp>. Paiva et al. found that HemoCue<sup>&#174; </sup>was more appropriate for capillary compared to venous blood samples <abbrgrp><abbr bid="B32">32</abbr></abbrgrp>. However, there was within-subject variability of capillary blood hemoglobin values that might explain the unreliability of the method, and it has been shown that two capillary samples taken from different fingers of the same subjects had hemoglobin concentrations differing by/ more than two g/dL using the HemoCue<sup>&#174; </sup><abbrgrp><abbr bid="B6">6</abbr></abbrgrp>.</p>
<p>In spite of the non-acceptable agreement of HemoCue<sup>&#174; </sup>with automated hematology analyzer in this study, the HemoCue<sup>&#174; </sup>is however simple to use, need minimum training, cheap, and gives an immediate result. Furthermore, it is useful in clinical and epidemiological settings where finger puncture allows capillary blood sampling as an easy technique which is less resource-intensive than vein puncture, and is more acceptable to patients and the community.</p>
</sec>
<sec><st><p>Conclusion</p></st>
<p>Hemoglobin concentration assessment by HemoCue<sup>&#174; </sup>using either venous or capillary blood samples does not have an acceptable agreement with automated hematology analyzer. Therefore this study does not recommend this device for hemoglobin concentration assessment among pregnant women.</p>
</sec>
<sec><st><p>Competing interests</p></st>
<p>The authors declare that they have no competing interests.</p>
</sec>
<sec><st><p>Authors' contributions</p></st>
<p>IA and MIY designed the study and participated in the statistical analyses, SA and MH conducted the clinical work. MH conducted the lab work. All the authors approved the draft and the final paper.</p>
</sec>
</bdy>
<bm>
<ack>
<sec><st><p>Acknowledgements</p></st>
<p>This work was funded by University of Khartoum, Khartoum, Sudan.</p>
</sec>
</ack>
<refgrp><bibl id="B1"><aug><au><cnm>WHO</cnm></au></aug><source>The world health report 2002: reducing risks, promoting healthy life. Geneva</source><pubdate>2002</pubdate></bibl><bibl id="B2"><title><p>Worldwide prevalence of anemia in preschool aged children, pregnant women and non-pregnant women of reproductive age</p></title><aug><au><snm>McLean</snm><fnm>E</fnm></au><au><snm>Egli</snm><fnm>I</fnm></au><au><snm>Cogswell</snm><fnm>M</fnm></au><au><snm>Benoist</snm><fnm>Bd</fnm></au><au><snm>Wojdyla</snm><fnm>D</fnm></au></aug><source>Nutritional anemia</source><publisher>Basel: Sight and Life Press</publisher><editor>Kraemer K, Zimmermann MB</editor><pubdate>2007</pubdate><fpage>1</fpage><lpage>12</lpage></bibl><bibl id="B3"><title><p>Perspective of the epidemiology of malaria and anaemia and their impact on maternal and perinatal outcomes in Sudan</p></title><aug><au><snm>Adam</snm><fnm>I</fnm></au><au><snm>Elhassan</snm><fnm>EM</fnm></au><au><snm>Haggaz</snm><fnm>AE</fnm></au><au><snm>Ali</snm><fnm>AA</fnm></au><au><snm>Adam</snm><fnm>GKA</fnm></au></aug><source>J Infect Dev Ctries</source><pubdate>2011</pubdate><volume>5</volume><fpage>83</fpage><lpage>87</lpage><xrefbib><pubid idtype="pmpid" link="fulltext">21389586</pubid></xrefbib></bibl><bibl id="B4"><title><p>Anaemia and stillbirth in Kassala Hospital, Eastern Sudan</p></title><aug><au><snm>Ali</snm><fnm>AA</fnm></au><au><snm>Adam</snm><fnm>I</fnm></au></aug><source>J Trop Pediatr</source><pubdate>2011</pubdate><volume>57</volume><fpage>62</fpage><lpage>64</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1093/tropej/fmq029</pubid><pubid idtype="pmpid">20427426</pubid></pubidlist></xrefbib></bibl><bibl id="B5"><title><p>Anaemia and low birth weight in Medani</p></title><aug><au><snm>Elhassan</snm><fnm>EM</fnm></au><au><snm>Abbaker</snm><fnm>AO</fnm></au><au><snm>Haggaz</snm><fnm>AD</fnm></au><au><snm>Abubaker</snm><fnm>MS</fnm></au><au><snm>Adam</snm><fnm>I</fnm></au></aug><source>Hospital Sudan. BMC Res Notes</source><pubdate>2010</pubdate><volume>3</volume><fpage>181</fpage></bibl><bibl id="B6"><title><p>Precision, accuracy, and reliability of hemoglobin assessment with use of capillary blood</p></title><aug><au><snm>Morris</snm><fnm>SS</fnm></au><au><snm>Ruel</snm><fnm>MT</fnm></au><au><snm>Cohen</snm><fnm>RJ</fnm></au><au><snm>Dewey</snm><fnm>KG</fnm></au><au><snm>de la Briere</snm><fnm>B</fnm></au><au><snm>Hassan</snm><fnm>MN</fnm></au></aug><source>Am J Clin Nutr</source><pubdate>1999</pubdate><volume>69</volume><fpage>1243</fpage><lpage>1248</lpage><xrefbib><pubid idtype="pmpid" link="fulltext">10357746</pubid></xrefbib></bibl><bibl id="B7"><title><p>The HemoCue<sup>&#174;</sup>, a point of care B-hemoglobin photometer, measures hemoglobin concentrations accurately when mixed in vitro with canine plasma and three hemoglobin-based oxygen carriers (HBOC)</p></title><aug><au><snm>Jahr</snm><fnm>JS</fnm></au><au><snm>Lurie</snm><fnm>F</fnm></au><au><snm>Driessen</snm><fnm>B</fnm></au><au><snm>Davis</snm><fnm>JA</fnm></au><au><snm>Gosselin</snm><fnm>R</fnm></au><au><snm>Gunther</snm><fnm>RA</fnm></au></aug><source>Can J Anesth</source><pubdate>2002</pubdate><volume>49</volume><fpage>243</fpage><lpage>248</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1007/BF03020522</pubid><pubid idtype="pmpid">11861341</pubid></pubidlist></xrefbib></bibl><bibl id="B8"><title><p>Evaluation of a new system for hemoglobin measurement</p></title><aug><au><snm>Bridges</snm><fnm>N</fnm></au><au><snm>Parvin</snm><fnm>RM</fnm></au><au><snm>Van Assendelft</snm><fnm>OW</fnm></au></aug><source>Am Clin Products Rev</source><pubdate>1987</pubdate><volume>6</volume><fpage>22</fpage><lpage>25</lpage></bibl><bibl id="B9"><title><p>Evaluation of HemoCue, a new device for determining hemoglobin</p></title><aug><au><snm>Von Schenck</snm><fnm>H</fnm></au><au><snm>Falkensson</snm><fnm>M</fnm></au><au><snm>Lundberg</snm><fnm>B</fnm></au></aug><source>Clin Chem</source><pubdate>1986</pubdate><volume>32</volume><fpage>526</fpage><lpage>529</lpage><xrefbib><pubid idtype="pmpid" link="fulltext">3948400</pubid></xrefbib></bibl><bibl id="B10"><title><p>Precision, accuracy, and reliability of hemoglobin assessment with use of capillary blood</p></title><aug><au><snm>Morris</snm><fnm>SS</fnm></au><au><snm>Ruel</snm><fnm>MT</fnm></au><au><snm>Cohen</snm><fnm>RJ</fnm></au><au><snm>Dewey</snm><fnm>KG</fnm></au><au><snm>de la Briere</snm><fnm>B</fnm></au><au><snm>Hassan</snm><fnm>MN</fnm></au></aug><source>Am J Clin Nutr</source><pubdate>1999</pubdate><volume>69</volume><fpage>1243</fpage><lpage>1248</lpage><xrefbib><pubid idtype="pmpid" link="fulltext">10357746</pubid></xrefbib></bibl><bibl id="B11"><title><p>A clinical evaluation of the HemoCue hemoglobinometer using capillary, venous and arterial samples</p></title><aug><au><snm>Chen</snm><fnm>PP</fnm></au><au><snm>Short</snm><fnm>TG</fnm></au><au><snm>Leung</snm><fnm>DH</fnm></au><au><snm>Oh</snm><fnm>TE</fnm></au></aug><source>Anaesth Intensive Care</source><pubdate>1992</pubdate><volume>20</volume><fpage>497</fpage><lpage>500</lpage><xrefbib><pubid idtype="pmpid">1463181</pubid></xrefbib></bibl><bibl id="B12"><title><p>HemoCue, an accurate bedside method of hemoglobin measurement?</p></title><aug><au><snm>Rippmann</snm><fnm>CE</fnm></au><au><snm>Nett</snm><fnm>PC</fnm></au><au><snm>Popovic</snm><fnm>D</fnm></au><au><snm>Seifert</snm><fnm>B</fnm></au><au><snm>Pasch</snm><fnm>T</fnm></au><au><snm>Spahn</snm><fnm>DR</fnm></au></aug><source>J Clin Monit</source><pubdate>1997</pubdate><volume>13</volume><fpage>373</fpage><lpage>377</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1023/A:1007451611748</pubid><pubid idtype="pmpid">9495289</pubid></pubidlist></xrefbib></bibl><bibl id="B13"><title><p>Evaluation of portable hemoglobinometer in general practice</p></title><aug><au><snm>Neville</snm><fnm>RG</fnm></au></aug><source>Br Med J (Clin Res Ed)</source><pubdate>1987</pubdate><volume>294</volume><fpage>1263</fpage><lpage>1265</lpage><xrefbib><pubid idtype="doi">10.1136/bmj.294.6582.1263</pubid></xrefbib></bibl><bibl id="B14"><title><p>Evaluation of a portable hemoglobin photometer in pregnant women in a high altitude area: a pilot study</p></title><aug><au><snm>Zhou</snm><fnm>X</fnm></au><au><snm>Yan</snm><fnm>H</fnm></au><au><snm>Xing</snm><fnm>Y</fnm></au><au><snm>Dang</snm><fnm>S</fnm></au><au><snm>Zhuoma</snm><fnm>B</fnm></au><au><snm>Wang</snm><fnm>D</fnm></au></aug><source>BMC Public Health</source><pubdate>2009</pubdate><volume>9</volume><fpage>228</fpage><xrefbib><pubidlist><pubid idtype="doi">10.1186/1471-2458-9-228</pubid><pubid idtype="pmcid">2717084</pubid><pubid idtype="pmpid" link="fulltext">19591672</pubid></pubidlist></xrefbib></bibl><bibl id="B15"><title><p>Diagnosing anaemia in pregnancy in rural clinics: assessing the potential of the Haemoglobin Colour Scale</p></title><aug><au><snm>van den Broek</snm><fnm>NR</fnm></au><au><snm>Ntonya</snm><fnm>C</fnm></au><au><snm>Mhango</snm><fnm>E</fnm></au><au><snm>White</snm><fnm>SA</fnm></au></aug><source>Bull World Health Organ</source><pubdate>1999</pubdate><volume>77</volume><fpage>15</fpage><lpage>21</lpage><xrefbib><pubidlist><pubid idtype="pmcid">2557566</pubid><pubid idtype="pmpid">10063656</pubid></pubidlist></xrefbib></bibl><bibl id="B16"><title><p>HemoCue<sup>&#174; </sup>Blood Hemoglobin Photometer Operating Manual</p></title><url>http://www.HemoCue&#174;.com/files/900138_B.pdf</url></bibl><bibl id="B17"><title><p>Performance Evaluation of the Sysmex KX-21 [TM] Automated Hematology Analyzer</p></title><aug><au><snm>Gamperling</snm><fnm>N</fnm></au><au><snm>Mast</snm><fnm>B</fnm></au><au><snm>Hagbloom</snm><fnm>R</fnm></au><au><snm>Houwen</snm><fnm>B</fnm></au></aug><source>Sysmex J Int</source><pubdate>1998</pubdate><volume>8</volume><fpage>96</fpage><lpage>101</lpage></bibl><bibl id="B18"><title><p>An evaluation of the Sysmex KX-21 automated hematology analyzer</p></title><aug><au><cnm>United Kingdom National External Quality Assessment Scheme for Haematology (UK NEQAS (H)), WGHV, Watford, WD1 8FJ, UK</cnm></au></aug><source>Sysmex J Int</source><pubdate>1998</pubdate><volume>8</volume><fpage>102</fpage><lpage>109</lpage></bibl><bibl id="B19"><title><p>Sysmex KX-21N Operator's Manual</p></title><source>Sysmex Corporation</source><pubdate>2006</pubdate></bibl><bibl id="B20"><title><p>Statistical methods for assessing agreement between two methods of clinical measurement</p></title><aug><au><snm>Bland</snm><fnm>JM</fnm></au><au><snm>Altman</snm><fnm>DG</fnm></au></aug><source>Lancet</source><pubdate>1986</pubdate><volume>1</volume><fpage>307</fpage><lpage>310</lpage><xrefbib><pubid idtype="pmpid">2868172</pubid></xrefbib></bibl><bibl id="B21"><title><p>Determination of capillary hemoglobin levels using the HemoCue system in intensive care patients</p></title><aug><au><snm>Seguin</snm><fnm>P</fnm></au><au><snm>Kleiber</snm><fnm>A</fnm></au><au><snm>Chanavaz</snm><fnm>C</fnm></au><au><snm>Morcet</snm><fnm>J</fnm></au><au><snm>Mall&#233;dant</snm><fnm>Y</fnm></au></aug><source>J Crit Care</source><pubdate>2011</pubdate><volume>26</volume><fpage>423</fpage><lpage>427</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1016/j.jcrc.2010.08.012</pubid><pubid idtype="pmpid" link="fulltext">21036530</pubid></pubidlist></xrefbib></bibl><bibl id="B22"><title><p>HemoCue, an accurate bedside method of hemoglobin measurement?</p></title><aug><au><snm>Rippmann</snm><fnm>CE</fnm></au><au><snm>Nett</snm><fnm>PC</fnm></au><au><snm>Popovic</snm><fnm>D</fnm></au><au><snm>Seifert</snm><fnm>B</fnm></au><au><snm>Pasch</snm><fnm>T</fnm></au><au><snm>Spahn</snm><fnm>DR</fnm></au></aug><source>J Clin Monit</source><pubdate>1997</pubdate><volume>13</volume><fpage>373</fpage><lpage>377</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1023/A:1007451611748</pubid><pubid idtype="pmpid">9495289</pubid></pubidlist></xrefbib></bibl><bibl id="B23"><title><p>Hemoglobin estimation by the HemoCue<sup>&#174; </sup>portable hemoglobin photometer in a resource poor setting</p></title><aug><au><snm>Nkrumah</snm><fnm>B</fnm></au><au><snm>Nguah</snm><fnm>SB</fnm></au><au><snm>Sarpong</snm><fnm>N</fnm></au><au><snm>Dekker</snm><fnm>D</fnm></au><au><snm>Idriss</snm><fnm>A</fnm></au><au><snm>May</snm><fnm>J</fnm></au><au><snm>Adu-Sarkodie</snm><fnm>Y</fnm></au></aug><source>BMC Clinical Pathology</source><pubdate>2011</pubdate><volume>11</volume><fpage>5</fpage><xrefbib><pubidlist><pubid idtype="doi">10.1186/1472-6890-11-5</pubid><pubid idtype="pmcid">3095531</pubid><pubid idtype="pmpid" link="fulltext">21510885</pubid></pubidlist></xrefbib></bibl><bibl id="B24"><title><p>Evaluation of "HemoCue", a new device for determining hemoglobin</p></title><aug><au><snm>Von Schenck</snm><fnm>H</fnm></au><au><snm>Falkensson</snm><fnm>M</fnm></au><au><snm>Lundberg</snm><fnm>B</fnm></au></aug><source>Clin Chem</source><pubdate>1986</pubdate><volume>32</volume><fpage>526</fpage><lpage>529</lpage><xrefbib><pubid idtype="pmpid" link="fulltext">3948400</pubid></xrefbib></bibl><bibl id="B25"><title><p>Reliability of HemoCue in patients with gastrointestinal bleeding</p></title><aug><au><snm>Van de Louw</snm><fnm>A</fnm></au><au><snm>Lasserre</snm><fnm>N</fnm></au><au><snm>Drouhin</snm><fnm>F</fnm></au><au><snm>Thierry</snm><fnm>S</fnm></au><au><snm>Lecuyer</snm><fnm>L</fnm></au><au><snm>Caen</snm><fnm>D</fnm></au><au><snm>Tenaillon</snm><fnm>A</fnm></au></aug><source>Intens Care Med</source><pubdate>2007</pubdate><volume>33</volume><fpage>355</fpage><lpage>358</lpage><xrefbib><pubid idtype="doi">10.1007/s00134-006-0461-6</pubid></xrefbib></bibl><bibl id="B26"><title><p>HemoCue, an Accurate Bedside Method of Hemoglobin Measurement?</p></title><aug><au><snm>Rippmann</snm><fnm>CE</fnm></au><au><snm>Nett</snm><fnm>PC</fnm></au><au><snm>Popovic</snm><fnm>D</fnm></au><au><snm>Seifert</snm><fnm>B</fnm></au><au><snm>Pasch</snm><fnm>T</fnm></au><au><snm>Spahn</snm><fnm>DR</fnm></au></aug><source>J Clin Monit Comput</source><pubdate>1997</pubdate><volume>13</volume><fpage>373</fpage><lpage>377</lpage></bibl><bibl id="B27"><title><p>Evaluation of portable haemoglobinometer in general practice</p></title><aug><au><snm>Neville</snm><fnm>RG</fnm></au></aug><source>BMJ (Clinical research ed)</source><pubdate>1987</pubdate><volume>294</volume><fpage>1263</fpage><lpage>1265</lpage><xrefbib><pubid idtype="doi">10.1136/bmj.294.6582.1263</pubid></xrefbib></bibl><bibl id="B28"><title><p>Evaluation of the HemoCue compared with the Coulter STKS for measurement of neonatal haemoglobin</p></title><aug><au><snm>Rechner</snm><fnm>IJ</fnm></au><au><snm>Twigg</snm><fnm>A</fnm></au><au><snm>Davies</snm><fnm>AF</fnm></au><au><snm>Imong</snm><fnm>S</fnm></au></aug><source>Arch Dis Child Fetal Neonatal</source><pubdate>2002</pubdate><volume>86</volume><fpage>188</fpage><lpage>189</lpage><xrefbib><pubid idtype="doi">10.1136/fn.86.3.F188</pubid></xrefbib></bibl><bibl id="B29"><title><p>Evaluation of the HemoCue for measuring intra operative haemoglobin concentrations: a comparison with the Coulter Max M<sup>&#174;</sup></p></title><aug><au><snm>Lardi</snm><fnm>AM</fnm></au><au><snm>Hirst</snm><fnm>C</fnm></au><au><snm>Mortimer</snm><fnm>AJ</fnm></au><au><snm>McCollum</snm><fnm>CN</fnm></au></aug><source>Anaesthesia</source><pubdate>1998</pubdate><volume>53</volume><fpage>349</fpage><lpage>352</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1046/j.1365-2044.1998.00328.x</pubid><pubid idtype="pmpid" link="fulltext">9613300</pubid></pubidlist></xrefbib></bibl><bibl id="B30"><title><p>Estimating the prevalence of anaemia: a comparison of three methods</p></title><aug><au><snm>Sari</snm><fnm>M</fnm></au><au><snm>dePee</snm><fnm>S</fnm></au><au><snm>Martini</snm><fnm>E</fnm></au><au><snm>Herman</snm><fnm>S</fnm></au><au><snm>Bloem</snm><fnm>MW</fnm></au><au><snm>Yip</snm><fnm>R</fnm></au></aug><source>Bull World Health Org</source><pubdate>2001</pubdate><volume>79</volume><fpage>506</fpage><lpage>511</lpage><xrefbib><pubidlist><pubid idtype="pmcid">2566437</pubid><pubid idtype="pmpid">11436471</pubid></pubidlist></xrefbib></bibl><bibl id="B31"><title><p>Hemoglobin screening in prospective blood donors: comparison of different blood samples and different quantitative methods</p></title><aug><au><snm>Radtke</snm><fnm>H</fnm></au><au><snm>Polat</snm><fnm>G</fnm></au><au><snm>Kalus</snm><fnm>U</fnm></au><au><snm>Salama</snm><fnm>A</fnm></au><au><snm>Kiesewetter</snm><fnm>H</fnm></au></aug><source>Transfus Aph Sci</source><pubdate>2005</pubdate><volume>33</volume><fpage>31</fpage><lpage>35</lpage><xrefbib><pubid idtype="doi">10.1016/j.transci.2004.11.004</pubid></xrefbib></bibl><bibl id="B32"><title><p>Comparison between the HemoCue<sup>&#174; </sup>and an automated counter for measuring hemoglobin</p></title><aug><au><snm>Paiva Adriana de</snm><fnm>A</fnm></au><au><snm>Rond&#243; Patr&#237;cia</snm><fnm>HC</fnm></au><au><snm>Silva Silmara</snm><fnm>S</fnm></au><au><snm>de</snm><fnm>B</fnm></au><au><snm>Latorre Maria do</snm><fnm>RDO</fnm></au><au><snm>de Paiva</snm><fnm>AA</fnm></au><au><snm>Rond&#243;</snm><fnm>PHC</fnm></au><au><snm>de Silva</snm><fnm>SSB</fnm></au><au><snm>Latorre</snm><fnm>MRDOdo</fnm></au></aug><source>Rev Saude Publica</source><pubdate>2004</pubdate><volume>38</volume><fpage>585</fpage><lpage>587</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1590/S0034-89102004000400017</pubid><pubid idtype="pmpid" link="fulltext">15311302</pubid></pubidlist></xrefbib></bibl></refgrp>
</bm>
</art>