Spectral- and time-domain optical coherence tomography measurements of macular thickness in young myopic eyes

Abstract

Purpose

To evaluate the variation in macular retinal thickness and volume in young Chinese myopic patients using time-domain optical coherence tomography (Stratus TD-OCT) and spectral-domain optical coherence tomography (Cirrus HD-OCT).

Methods

Ninety-two eyes of 92 myopic subjects were recruited in this study. Based upon spherical equivalence (SE), subjects were divided into two groups: the low to moderate myopia group (-0.5 D ≤ SE < -6.0 D), and the high myopia group (SE ≥ -6.0 D). Stratus TD-OCT and Cirrus HD-OCT were used to compare macular retinal thickness and volume between the two groups. Bland–Altman analysis and Pearson correlation were used to measure agreement between the two OCT systems.

Results

Average macular retinal thickness and total macular volume measured by Cirrus HD-OCT and Stratus TD-OCT of the low to moderate myopia group were 283.52 ± 12.14 μm and 245.38 ± 8.55 μm, respectively, and 10.08 ± 0.37 mm3 and 6.85 ± 0.26 mm3, respectively, and the high myopia groups were 269.58 ± 10.72 μm and 235.65 ± 7.54 μm, respectively, and 9.71 ± 0.36 mm3 and 6.52 ± 0.25 mm3, respectively. The measurements of the two OCTs showed that macular retinal thickness of the parafovea was significantly lower in the high myopia group compared with the low to moderate myopia group, except at the fovea (all P-values less than 0.001, except at the fovea). Using the Bland–Altman method and Pearson correlation, measurements of macular thickness in nine macular retinal subfields and total macular volumes showed good agreement between the two OCTs in myopic eyes (all P-values less than 0.001), with better agreement in the low to moderate myopia group than in the high myopia group.

Conclusions

The average macular retinal thickness of the fovea did not vary with myopia, while the total volume and retinal thickness of the parafovea were thinner with increasing myopia. There was good agreement between the two OCTs in myopic eyes in all macular subfields, and the Cirrus HD-OCT system provided thicker macular retinal thickness measurements than the Stratus TD-OCT system.

Virtual slides

The virtual slides for this article can be found here:http://www.diagnosticpathology.diagnomx.eu/vs/1683223414107652.

Background

In eastern Asia, myopia has reached epidemic proportions, and the prevalence of myopia has continued to increase[13]. High myopia is one of the commonest types of human ocular optical disorders, which can result in complications such as cataract, posterior scleral staphyloma, retinal detachment, choroidal neovascularization, and macular holes[48]. The retina can degenerate and become thinner in myopic eyes, especially at the posterior pole[9], so characterizing the associations between macular thickness of myopia and macular diseases has important clinical implications.

Optical coherence tomography (OCT) is a noninvasive, cross-sectional imaging technique that can measure macular thickness, and is highly reproducible[1012]. Apart from its diagnostic value in uveitis, OCT has enabled objective assessment of treatment response, provided predictive value for visual recovery and prognosis of uveitic entities, and detected toxic effects of the drug[13, 14]. The purpose of this study was to evaluate the variations in macular retinal thickness and volume in young Chinese myopic patients by using time-domain optical coherence tomography (Stratus TD-OCT) and spectral-domain optical coherence tomography (Cirrus HD-OCT), to identify the morphological features of the macula in macular diseases. These techniques could evaluate risk factors associated with myopia, which may assist in early diagnosis and monitoring of macular changes.

Material and methods

Subjects

Ninety-two eyes of 92 consecutive refractive surgery candidates with spherical equivalents ≥ -0.5 diopters (D) were recruited for the study. Liu et al. have reported[15] corneal astigmatism effects on the OCT measurements among high myopes, therefore corneal astigmatism of each subject was less than -1.0 D, especially in the high myopia group. The study has been approved by the Institutional Ethnic Committee (Institutional Review Board) of Shanghai Jiaotong University Affiliated Sixth People’s Hospital, and informed consent was obtained for each patient.

Each subject underwent a full ophthalmic examination, which included measurements of visual acuity, refraction, and intraocular pressure (IOP) by a noncontact tonometer. Axial length measurements were obtained in each eye with the IOL Master (Carl Zeiss Meditec, Inc, Dublin, CA, USA). Optic nerve head evaluation was performed with a 90-D lens. Macular retinal thickness and volume were measured by a Stratus TD-OCT and Cirrus HD-OCT (Carl Zeiss Meditec, Dublin, CA, USA).

Inclusion criteria were as follows: best corrected visual acuity (BCVA) of 20/20 or better, an IOP lower than 21 mm Hg in either eye, a healthy optic nerve head (ONH) without glaucomatous damage (i.e., no disc hemorrhage, notching, or thinning of the neural rim).

Those with a history of severe ocular trauma, intraocular or refractive surgery, or any ocular or neurological disease that could have affected the ONH or retinal nerve fiber layer (RNFL) were excluded from the study. Subjects with evidence of macular disease or peripapillary atrophy extending more than 1.73 mm from the center of the optic disc or with glaucoma or an IOP higher than 21 mmHg in either eye were also excluded. In addition, participants with a history of systemic diseases including hypertension and diabetes were excluded.

OCT imaging

After pupillary dilation to a minimum diameter of 5 mm, eyes of the subjects that satisfied the study criteria were scanned using the Cirrus HD-OCT system with software version 5.0 and Stratus TD-OCT system with software version 4.0. The mean retinal thickness and volume maps were determined for nine sectors in three concentric circles of diameters 1, 3, and 6 mm. The inner and outer rings were divided into four quadrants as follows: outer of temporal (To), outer of superior (So), outer of inferior (Io), outer of nasal (No), inner of temporal (Ti), inner of superior (Si), inner of inferior (Ii), and inner of nasal (Ni). The foveal thickness (F) was defined as the average thickness in the central 1 mm diameter. All the scans had signal strength of at least six and all measurements were taken by a single, trained examiner.

Statistical analysis

Statistical analyses were performed with commercially available software (SPSS ver. 12.0; SPSS Inc, Chicago, IL, USA). The total average and mean macular retinal thickness and macular volume were compared between the two groups with an independent t-test (except for the gender, which was compared by the chi-square test). Agreements between the two OCT systems were measured by Bland–Altman method and the Pearson correlation. A P-value less than 0.05 was considered statistically significant.

Results

Patient characteristics

Of ninety-two eyes, the mean age of the subjects was 22.89 ± 3.27-years-old, the average spherical equivalent (SE) was -5.64 ± 2.66 D (diopters) with a range of -0.38 D to -11.00 D. Based upon spherical equivalence (SE), subjects were divided into two groups: the low to moderate myopia group (-0.5 D ≤ SE < -6.0 D), and the high myopia group (SE ≥ -6.0 D). Forty-four eyes were classified as low to moderate myopia group with a mean SE of -3.34 ± 1.28 D, and 48 eyes were classified as high myopia group with an average SE of -7.75 ± 1.65 D. Characteristics of the two groups are listed in Table 1. No significant differences were found for age, sex, axial length, and spherical equivalence between the two groups.

Table 1 Characteristics of the LMM and HM groups x ¯ ± s

Topographic profile of macular retinal thickness among young myopic patients

From Table 2, the average macular retinal thickness and total macular volume measured by Cirrus HD-OCT and Stratus TD-OCT of the low to moderate myopia group were 283.52 ± 12.14 μm and 245.38 ± 8.55 μm, 10.08 ± 0.37 mm3 and 6.85 ± 0.26 mm3, respectively, and of the high myopia group were 269.58 ± 10.72 μm and 235.65 ± 7.54 μm, 9.71 ± 0.36 mm3 and 6.52 ± 0.25 mm3, respectively. Measurements using the two OCTs showed that the macular retinal thickness of the parafovea was significantly lower in the high myopia group compared with the low to moderate myopia group, except at the fovea (all P-values less than 0.001, except at the fovea).

Table 2 Comparison of macular thickness and volume as determined by optical coherence tomography (OCT)

Correlations between macular retinal thickness, macular volume, and axial length

The change of axial length is a characteristic of myopic eyes, so the morphological features of the macula may change with axial length. As shown in Table 3, the average macular retinal thickness of the parafovea and macular volume decreased with axial length, (all P-values less than 0.001, except at the fovea), especially in outer parafovea retinal subfields. These results demonstrated that with increase of axis length, macular thickness became thinner.

Table 3 Correlation analyses between axial length and retinal thickness/volume

Comparisons of OCT measurements of macular thickness and macular volume

Table 2 and Table 4 show that the Cirrus HD-OCT system provided thicker macular retinal thickness measurements than the Stratus TD-OCT system. The measurements of macular thickness in nine macular retinal subfields and total macular volume showed good agreement between the two OCTs in myopic eyes in all macular subfields, using the Bland–Altman method and Pearson correlation (all P-values less than 0.001), with a better agreement in the low to moderate myopia group than in the high myopia group (Figure 1).

Table 4 Bland-Altman analysis of macular thickness and volume values between Stratus TD-OCT and Cirrus HD-OCT
Figure 1
figure 1

Bland–Altman plots of average macular retinal thickness/macular volume values between Stratus TD-OCT and Cirrus HD-OCT. (A) Average macular retinal thickness in the low to moderate myopia group; (B) average macular retinal thickness in the high myopia group; (C) total macular volume in the low to moderate myopia group; (D) total average macular volume in the high myopia group. The difference (Cirrus HD-OCT-Stratus/TD-OCT) is plotted against the average of both measurements (Cirrus HD-OCT + Stratus TD-OCT/2) for all participants in each group. SD, standard deviation.

Discussion

Myopia is one of the most common causes of visual impairment and blindness, so early detection and management of degenerative eye diseases is of utmost importance. The results of the current study demonstrated that average macular retinal thickness of the fovea did not vary with myopia, while total volume and the retinal thickness of the parafovea were thinner with increased myopia. There was good agreement between the two OCTs in myopic eyes in all macular subfields, and the Cirrus HD-OCT system provided thicker macular retinal thickness measurements than the Stratus TD-OCT system.

Previous studies reported that Cirrus HD-OCT measurements resulted in a significantly thicker retinal thickness than the Stratus TD-OCT[1618]. We also found that Cirrus HD-OCT measurements resulted in thicker retinal thickness measurements than the Stratus TD-OCT in myopic eyes. Different technologies and segmentation algorithms of the two OCT systems may explain the thicker retinal thickness measurements observed in Cirrus HD-OCT. The most important difference is that the Cirrus segmentation identified the thickness of the retina from the retinal pigment epithelium (RPE) to the internal limiting membrane (ILM), while the Stratus segmentation identified retinal thickness from the inner segment/outer segment junction (IS/OS) to the ILM, thus Cirrus HD-OCT measurements resulted in an inherently thicker distance between the IS/OS and the RPE[19].

Our study showed that the two OCTs showed macular retinal thickness of the parafovea, especially in outer retinal subfields, that was significantly lower in the high myopia group compared with the low to moderate myopia group, except at the fovea (all P-values less than 0.001, except at the fovea). Other studies demonstrated that the average macular retinal thickness of the parafovea and total macular volume decreased with axial length (all P-values less than 0.001, except at the fovea). So with an increase of axis length, macular thickness becomes thinner, especially in the outer retinal subfields. Marcus et al. and Lim et al. also reported that the parafovea was thinner in myopic patients’ myopia[20]. This observation may be the result of the anatomical features of the retinal macular foveolar region. The foveolar is the thinnest area of the retina, which consists of the five innermost cell layers of the retina, including the outer plexiform layer, the outer nuclear layer, the external limiting membrane, the photoreceptor layer, and the retinal pigment epithelium. Previous histopathologic studies reported that the thickness of neurosensory retina (RNL), especially the inner nuclear layer, were thinner in several animal models of myopia[21]. The foveolar lacks the innermost retinal layers, and there is no inner nuclear, so the retinal thickness variation is not obvious between the high myopia group and moderate myopia groups. Therefore, the macular retinal thickness in the parafovea was thinner in the high myopia group than in the moderate myopia group, but no difference was found in the foveolar. Moreover, other pathological processes, such as choroid retinal degeneration, and choroid and RPE atrophy, could likewise play a role in changes of the retina.

The results of the current study characterized macular retinal thickness in young myopic patients. These characteristics can be used as an auxiliary diagnosis for making a distinction with other macular retinal related diseases. If the macular thickness in the foveolar is abnormally thin, pathological disease should be considered. However, if the changes occurred in the parafovea, other factors can first be eliminated, such as refractive factors, which could influence the measurements of macular thickness by the OCTs, especially among high myopia patients. In conclusion, when the macular thickness is abnormal, clinicians should consider the effect of the refractive factor combined with clinical manifestations to comprehensively evaluate its clinical significance.

References

  1. Bar Dayan Y, Levin A, Morad Y, Grotto I, Ben-David R, Goldberg A, Onn E, Avni I, Levi Y, Benyamini OG: The changing prevalence of myopia in young adults: a 13-year series population-based prevalence surveys. Invest Ophthalmol Vis Sci. 2005, 46 (8): 2760-2765. 10.1167/iovs.04-0260.

    Article  PubMed  Google Scholar 

  2. Wong TY, Foster PJ, Hee J, Ng TP, Tielsch JM, Chew SJ, Johnson GJ, Seah SK: Prevalence and risk factors for refractive errors in adult Chinese in Singapore. Invest Ophthalmol Vis Sci. 2000, 41: 2486-2494.

    PubMed  CAS  Google Scholar 

  3. Grosvenor T: Why is there an epidemic of myopia?. Clin Exp Optom. 2003, 86: 273-275. 10.1111/j.1444-0938.2003.tb03122.x.

    Article  PubMed  Google Scholar 

  4. Lim R, Mitchell P: Cumming RG Refractive associations with cataract: the blue mountains eye study. Invest Ophthalmol Vis Sci. 1999, 40 (12): 3021-3026.

    PubMed  CAS  Google Scholar 

  5. Mitchell P, Hourihan F, Sandbach J, Wang JJ: The relationship between glaucoma and myopia: The Blue Mountains Eye Study. Ophthalmology. 1999, 106: 2010-2015. 10.1016/S0161-6420(99)90416-5.

    Article  PubMed  CAS  Google Scholar 

  6. Stirpe M, Michels RG: Retinal detachment in highly myopic eyes due to macular holes and epiretinal traction. Retina. 1990, 10 (2): 113-114. 10.1097/00006982-199004000-00004.

    Article  PubMed  CAS  Google Scholar 

  7. Saw SM, Gazzard G, Shih-Yen EC, Chua WH: Myopia and associated pathological complications. Ophthalmic Physiol Opt. 2005, 25 (5): 381-391. 10.1111/j.1475-1313.2005.00298.x.

    Article  PubMed  Google Scholar 

  8. Wang FR: Current status and problems in the study of myopia. Zhong hua Yan Ke Za Zhi. 2003, 39 (6): 381-384.

    Google Scholar 

  9. Yanoff M, Fine BS: Ocular Pathology: A Text and Atlas. 1989, Philadelphia: JB Lippincott, 408-3

    Google Scholar 

  10. van Velthoven ME, Faber DJ, Verbraak FD, van Leeuwen TG, de Smet MD: Recent developments in optical coherence tomography for imaging the retina. Progr Retin Eye Res. 2007, 26: 57-77. 10.1016/j.preteyeres.2006.10.002.

    Article  Google Scholar 

  11. Chen TC, Cense B, Pierce MC, Nassif N, Park BH, Yun SH, White BR, Bouma BE, Tearney GJ, de Boer JF: Spectral domain optical coherence tomography: ultra-high speed, ultra-high resolution ophthalmic imaging. Arch Ophthalmol. 2005, 123: 1715-1720. 10.1001/archopht.123.12.1715.

    Article  PubMed  Google Scholar 

  12. Paunescu LA, Schuman JS, Price LL, Stark PC, Beaton S, Ishikawa H, Wollstein G, Fujimoto JG: Reproducibility of nerve fiber thickness, macular thickness, and optic nerve head measurements using Stratus OCT. Invest Ophthalmol Vis Sci. 2004, 45 (6): 1716-1724. 10.1167/iovs.03-0514.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Ma XY, Yan L, He L, He D, Lu H: Ocular fundus manifestation of two patients following long-term chloroquine therapy: a case report. Diagn Pathol. 2010, 5: 20-10.1186/1746-1596-5-20.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Onal S, Tugal-Tutkun I, Neri P, P Herbort C: Optical coherence tomography imaging in uveitis. Int Ophthalmol. 2013, 7: 9-

    Google Scholar 

  15. Liu L, Zou J, Huang H, Yang JG, Chen SR: The influence of corneal astigmatism on retinal nerve fiber layer thickness and optic nerve head parameter measurements by spectral-domain optical coherence tomography. Diagn Pathol. 2012, 7: 55-10.1186/1746-1596-7-55.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Giammaria D, Ioni A, Bartoli B, Cofini V, Pellegrini G, Giannotti B: Comparison of macular thickness measurements between time-domain and spectral-domain optical coherence tomographies in eyes with and without macular abnormalities. Retina. 2011, 31 (4): 707-716.

    PubMed  Google Scholar 

  17. Kakinoki M, Sawada O, Sawada T, Kawamura H, Ohji M: Comparison of macular thickness between Cirrus HD-OCT and Stratus OCT. Ophthalmic Surg Lasers Imaging. 2009, 40 (2): 135-140. 10.3928/15428877-20090301-09.

    Article  PubMed  Google Scholar 

  18. Han IC, Jaffe GJ: Comparison of spectral-and time-domain optical coherence tomography for retinal thickness measurements in healthy and diseased eyes. Am J Ophthalmol. 2009, 147 (5): 847-858. 10.1016/j.ajo.2008.11.019. 858.e1

    Article  PubMed  Google Scholar 

  19. Shin HJ, Cho BJ: Comparison of retinal nerve fiber layer thickness between Stratus and Spectralis OCT. Korean J Ophthalmol. 2011, 25 (3): 166-173. 10.3341/kjo.2011.25.3.166.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Lim MC, Hoh ST, Foster PJ, Lim TH, Chew SJ, Seah SK, Aung T: Use of optical coherence tomography to assess variations in macular retinal thickness in myopia. Invest Ophthalmol Vis Sci. 2005, 46 (3): 974-978. 10.1167/iovs.04-0828.

    Article  PubMed  Google Scholar 

  21. Zhong XW, Ge J, Chen XL, Nie HH, Huang J: Comparison of retinal morphology and ultrastructure in defocus-induced myopia and form-deprivation myopia in rhesus monkeys. Zhonghua Yan Ke Za Zhi. 2005, 41 (7): 625-630.

    PubMed  Google Scholar 

Download references

Acknowledgements

Supported by The Key Program for Fundamental Research of Shanghai Science and Technology commission, China. (NO.11JC1401202).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jun Zou.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

LL participated in the study design, reviewed the literature, collected the clinical data, and drafted the manuscript. JZ provided the conception and design of the study and reviewing the manuscript. L-lJ collected the clinical data and selected the material. J-gY took part in the study design and performed the statistical analysis. S-rC participated in collected the clinical data. All authors have read and approved the manuscript.

Authors’ original submitted files for images

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( https://creativecommons.org/publicdomain/zero/1.0/ ) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Cite this article

Liu, L., Zou, J., Jia, L. et al. Spectral- and time-domain optical coherence tomography measurements of macular thickness in young myopic eyes. Diagn Pathol 9, 38 (2014). https://doi.org/10.1186/1746-1596-9-38

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1746-1596-9-38

Keywords