Table 2 |
|||
|
Correlation, bias, and limits of agreement between hemoglobin level using HemoCue® and automated hematology analyzer |
|||
|
Comparison of methods |
Correlation Coefficient |
Bias ± SD (95%CI) |
Limits of agreement |
|
|
|||
|
HemoCue® (venous) vs. Hematology analyzer |
0.58 |
1.17 ± 1.57 (0.87; 1.47) |
-1.97 to 4.31 |
|
HemoCue® (capillary) vs. Hematology analyzer |
0.51 |
1.34 ± 1.85 (0.99; 1.69) |
-2.36 to 5.04 |
|
HemoCue® (venous) vs. HemoCue® (capillary) |
0.51 |
-0.17 ± 1.90 (-0.53; 0.19) |
-3.97 to 3.63 |
|
|
|||
|
Adam et al. Diagnostic Pathology 2012 7:30 doi:10.1186/1746-1596-7-30 |
|||